9.4 MyMode "Freeze position"

(if the knee joint is permanently locked)
Switching process
In this MyMode, the joint is locked in its current position and cannot be moved
in the flexion nor the extension direction.
1) Flex or extend the knee joint to the angle where it should be locked.
2) On the screen of the device, tap the corresponding symbol to activate the
3) The orthosis is locked in the flexion and extension directions.
Deactivate the lock by selecting a different MyMode/basic mode or by turning
the component off and then on again.

9.5 MyMode "User defined"

The MyMode "User defined" is intended for specific motion patterns or pos­
tures (e.g. cycling ...). This MyMode can be configured individually by the
O&P professional using the adjustment app.
Settings can be adjusted using the Cockpit app (see Page 84).

10 Additional operating states (modes)

10.1 Empty battery mode

Beeps and vibration signals are emitted if the available battery charge level drops to 5% (see
Page 94). The settings are changed to the safety mode parameters during this time. Then the
product is switched off.
INFORMATION
After disconnecting the battery charger, the orthosis is in the same state it was in before con­
necting the battery charger. For example, if the orthosis was switched off before connecting the
battery charger, it will also be switched off after disconnecting the battery charger. A descend­
ing sequence of beeps is emitted when the battery charger is disconnected while the orthosis is
switched off
.

10.2 Mode for charging the product

The product is non-functional during charging.
To switch to basic mode, the battery charger for the product must be disconnected after the bat­
tery is charged.

10.3 Safety mode

The product automatically switches to safety mode if a critical system fault occurs (e.g. failure of a
sensor signal). Safety mode remains in effect until the error has been rectified.
Flexion resistance preconfigured by the O&P professional is activated in safety mode, corres­
ponding at least to the stance phase resistance. The extension resistance is low and cannot be
88
lock.
→ A confirmation signal sounds to indicate activation of the lock.