changed. This makes limited walking possible for the user and allows the user to sit down, even
though the sensor system is not active.
The switch to safety mode is indicated by the red
and vibration signals immediately prior to switching (see Page 94).
Safety mode can be disabled by switching the product off/on (see Page 85) and by connecting
then disconnecting the battery charger. If the product switches into safety mode again, this
means a permanent error exists. The product must be inspected immediately by an O&P profes­
sional.

10.4 Overheating mode

When the hydraulic unit overheats due to uninterrupted, increased activity (e.g. extended walking
downhill), the flexion resistance is increased along with the rising temperature in order to counter­
act the overheating. When the hydraulic unit cools down, the product switches back to the set­
tings that existed before the overheating mode.
Overheating mode is not activated in the MyMode "User defined".
Overheating mode is indicated by a long vibration every 5 seconds and additionally by the orange
symbol lighting up.
The following functions are deactivated in overheating mode:
Sitting function
Switching to a MyMode
Changing the orthosis settings

11 Cleaning

1) Clean the product with a damp cloth and mild soap when needed.
2) Dry the product with a lint-free cloth and allow it to air dry fully.
Textile component
1) Remove the textile component from the product.
2) Fasten all hook-and-loop closures.
3) Hand wash the textile component in warm water at 30 °C/86 °F with a standard mild deter­
gent. Rinse thoroughly.
4) Allow to air dry. Do not expose to direct heat sources (e.g. sunlight, stove or radiator).

12 Maintenance

Regular service inspections (maintenance) are mandatory in the interest of your own safety and in
order to maintain operating reliability and protect the warranty. These service inspections include
an inspection of the sensors and replacement of wear and tear parts.
If maintenance is due, this is indicated after the service interval has expired by the
the control panel of the joint unit lighting up after connecting the battery charger. For further
information, please contact your O&P professional.
To have a service inspection carried out, please submit the orthosis as well as the battery charger
and power supply unit to the O&P professional.
17KO1=* C-Brace joint unit
The manufacturer prescribes a service inspection at intervals of 2 years for the product.
Additional orthosis components used, e.g. joints
Observe the service intervals and maintenance instructions for all orthosis components used.

13 Legal information

All legal conditions are subject to the respective national laws of the country of use and may vary
accordingly.