The product must be inspected immediately by an O&P professional.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Observe the warnings/error signals (see Page 94).
Do not continue using the product after the safety signal has been emitted.
The product must be inspected by an authorised Ottobock Service Centre. In this case, the
contact person is your prosthetist.
CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
The warnings/error signals (see Page 94) have to be observed.
Particular caution is necessary when using a bicycle without a freewheel (with a fixed gear).

4.11 Information on the use of a mobile device with the cockpit app

CAUTION
Improper use of the device
Falling due to altered damping behaviour as a result of unexpected switching into MyMode.
Make sure you have been instructed in the proper use of the device with the cockpit app.
CAUTION
Independently applied changes or modifications made to the device
Falling due to altered damping behaviour as a result of unexpected switching into MyMode.
Do not make any independent changes to the hardware of the device.
Do not make any independent changes to the software/firmware of the device which are not
included in the update function of the software/firmware.
CAUTION
Improper mode switching with the device
Falling due to unexpected product behaviour because of changed damping behaviour.
Ensure that you stand securely during all switching processes.
Verify the changed damping characteristics after switching and observe the feedback from
the acoustic signal emitter (beeper) and the display on the device.
Switch back to basic mode once the activities in MyMode have been completed.
NOTICE
Destruction of the device due to falling or penetration of water
Malfunction of the device.
If necessary, let the device dry at room temperature (at least 1 day).
If it is no longer be possible to switch back from a MyMode to basic mode, the component
can only be switched back to basic mode by turning it off/on (see Page 85).