6800 Enclosure; Location; Pipe Mount Instructions - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

The 6700 enclosure supports battery-backed operation for the NGC. The
XFC/XRC electronic board provides a battery charger/regulator for the system.

3.29.3 6800 Enclosure

The 6800 enclosure can accommodate the following :

3.29.4 Location

Mount the enclosure on a nearby wall, panel or pole. Make sure that the approved
conduit can be installed between the power supply's enclosure and the NGC.
Avoid obstructions.

3.29.5 Pipe Mount Instructions

Enclosure mounting brackets and fastening hardware are supplied with the unit.
The customer must provide 2" pipe of suitable length (see
If a charging source such as a solar panel is desired, this procedure may be
adapted to mount the solar panel on the upper portion of the pipe.
The instructions assume the mounting pipe has been previously installed. If not,
refer to installation sections previously in this chapter for either free-standing pipe
installation or pipe saddle installation.
1) When the unit is received, unpack and inspect all components for evidence
2) Using instructions supplied with the mounting kit, attach the bracket to the back
of enclosure unit.
3) Position the unit on the 2" mounting pipe and secure in place with two U-bolts,
split washers, flat washers and two bolts (see
Figure
Page 3–56
1 each 26/30 AH battery
Communications kit
Solar Panel Power Option
115/230 VAC UPS Power Option (24 VDC systems only)
of damage. Report damage to the shipping carrier and to the Totalflow
service department.
3-61).
Figure
3-59).
Figure
3-59,
Figure 3-60
2101510 Rev. AG
and

Advertisement