6.3.11.1
These alarms are indicative of a change to the CV Percent of sufficient
percentage to activate the alarm. This alarm will discontinue a scheduled
calibration and will need to be disabled prior to calibrating the unit.
6.3.11.2
1) On the Analyzer Operation screen, click Hold under Next Mode. When the unit
completes the current cycle and enters hold, continue to the next step.
2) Verify the calibration blend concentrations to calibration blend concentrations
listed on the Calibration Setup screen. If the errors exist, make corrections and
send the setup when complete.
3) Under Stream Setup, Alarm Definitions, locate the Calibration CV Percent Error
Alarm and set Alarm Enable to No. Send the change. Repeat for any additional
streams with this alarm.
4) Following the
ensuring that the Next Mode is set to Hold.
5) When the unit enters hold, select Peak Find.
6) Verify that the peaks are correctly labeled and integrated. If the peaks are
correctly labeled and integrated, return the unit to operation.
7) Reset the Alarm Enable to Yes. Verify that the alarm threshold is a valid
configuration.
8) Return the unit to regular operation.

6.3.12 Calibration RF Percent Error Alarm

If the Response Factor (RF) Percent Error alarm is in fault status, the following
procedure will step the user through the troubleshooting process. On occasion,
these instructions may detour the user to other procedures, and, when complete,
the user should return to these procedures to continue.
6.3.12.1
These alarms are indicative of a change to the response factor of sufficient
percentage to activate the alarm. This alarm will discontinue a scheduled
calibration and will need to be disabled prior to calibrating the unit.
6.3.12.2
1) Verify the calibration blend concentrations to calibration blend concentrations
listed on the Calibration Setup screen. If the errors exist, make corrections and
send the setup when complete.
2) On the Analyzer Operation screen, click Hold under Next Mode. When the unit
completes the current cycle and enters hold, continue to the next step.
3) Under Stream Setup, Alarm Definitions, locate the RF Percent Error Alarm and
set Alarm Enable to No. Send the change. Repeat for any additional streams
with this alarm.
4) When the unit enters hold, select Peak Find. Select Run Auto PF.
Page 6–14
Description
Instructions
Calibrating the NGC
The information provided for troubleshooting this alarm is only
intended to cover basic steps that can be performed in the
field. On occasion, additional troubleshooting steps may be
provided by Totalflow technical support in an effort to reduce
down time. Additionally, it may be desirable to return a
module to Totalflow for comprehensive testing and/or repair.
Description
Instructions
instructions in Startup, perform a calibration