3.32.1 Instructions

1) If configured, the optional equipment unit should contain an installed AC
2) Remove the necessary plugs from the side of the enclosure to install the rigid
conduit.
3) Pipe the conduit and associated AC wiring into the enclosure.
4) Using wiring instructions in
VAC hot and neutral or the two hot wires for 230 VAC to TB1 terminal 5A and
terminal 6A on the power supply.
5) Pipe the conduit and associated DC wiring from the NGC into the power supply
enclosure. See
6) Remove the J1 connector from the NGC termination panel. Using the wiring
instructions in
from F1 terminal B to J1 connector (+) pin and connect the ground wire TB1
terminal 1A to J1 connector (-) pin. Do not re-insert the J1 connector to the
termination board.
7) Go to
DIGITAL I/O
POWER
J2
(-)
(+) (-) (+) (-) (+) (-) (+)
(-)
1
2
1
2
(+)
INPUTS
OUTPUTS
5VDC
D2
J7
(+)
(-)
J9
TERM
1
3
NO TERM
SERIAL PORT 1
J8
1
2
3
4
5
6
7
8
9
J5
J6
USB HOST
USB CLIENT
Termination Panel
NGC8200

Figure 3-71 AC/DC Converter Wiring Instructions

2101510 Rev. AG
The AC/DC power supply may be approved for classified
hazardous locations or potentially explosive atmospheres.
Verify the rating listed on the unit tag, and install per the
referenced control drawing. Be sure to follow the requirements
of national and local codes when installing the power supply.
Installation must be performed by person(s) qualified for the
type and area of installation, according to national and local
codes.
power supply. The Optional equipment enclosure should be installed using
instructions shown previously in this chapter.
Review the Grounding the NGC section in
Description, before making power connections.
Table 2–4
Figure
3-71, make field connections from the power supply wire
DC Power Installation
CAUTION: Pay careful attention to
polarity when connecting power to the
NGC8200
J1
Customer Supplied
Wiring From Power
Supply
Figure
3-71, make the field connections from the 115
in Chapter 1 for wire sizes.
later in this chapter.
If this unit is installed in an office or lab environment where a dedicated earth
ground is not available, jumper AC GND to Output (-). DO NOT install this
jumper in field installations where a dedicated earth grounding is employed.
Phoenix Power Supply
System
To Dedicated
Earth GND
100-240 VAC
Power Source
Page 3–67