Stream Sequence Error Alarm; Calibration Cv Percent Error Alarm - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

2) Verify the calibration blend concentrations to calibration blend concentrations
listed on the Calibration Setup screen. If an error exists, make corrections and
send the setup when complete.
3) Under Stream Setup, Alarm Definitions, locate the calibration un-normalized error
alarm and set alarm enable to No. Send change. Repeat for any additional
streams with this alarm.
4) Following the
ensuring that the Next Mode is set to Hold.
5) When the unit enters hold, select Peak Find.
6) Verify that the peaks are correctly labeled and integrated. If the peaks are
correctly labeled and integrated, return the unit to operation; otherwise, continue
to the next step.
7) Select Peak Find from the Analyzer Operation screen. Ensure that Automatic is
check and then select Run Auto PF. This procedure will require approximately
45 minutes.
8) When the unit enters hold, verify that the peaks are correctly labeled and
integrated. If the peaks are correctly labeled and integrated, return the unit to
operation; otherwise, contact Totalflow technical support.
9) Reset the Alarm Enable to Yes. Verify that the alarm threshold is a valid
configuration. Typically, the un-normalized total should be within 6.50% (between
99.5 and 100.5).
10) Return the unit to regular operation.

6.3.10 Stream Sequence Error Alarm

If the Stream Sequence Error alarm is in fault status, the following procedure will
step the user through the troubleshooting process. On occasion, these instructions
may detour the user to other procedures, and, when complete, they should return
to these procedures to continue.
6.3.10.1
These alarms are indicative of a synchronization problem following a manual data
post process in factory mode.
6.3.10.2
1) On the Analyzer Operation screen, click Hold under Next Mode. When the unit
completes the current cycle and enters hold, continue to the next step.
2) Following the
start.

6.3.11 Calibration CV Percent Error Alarm

If the Calibration CV Percent Error alarm is in fault status, the following procedure
will step the user through the troubleshooting process. On occasion, these
instructions may detour the user to other procedures, and, when complete, they
should return to these procedures to continue.
2101510 Rev. AG
Calibrating the NGC
The information provided for troubleshooting this alarm is only
intended to cover basic steps that can be performed in the
field. On occasion, additional troubleshooting steps may be
provided by Totalflow technical support in an effort to reduce
down time. Additionally, it may be desirable to return a
module to Totalflow for comprehensive testing and/or repair.
Description
Instructions
Reset Procedures
instructions in Startup, perform a calibration
instructions in Maintenance, perform a warm
Page 6–13

Advertisement

Table of Contents
loading

Table of Contents