3.28 Sealing Environmental Enclosure

3.28.1 Customer Supplied Materials

Aerosol insulating foam

3.28.2 Instructions

1) When all sample and vent connections are complete, apply aerosol

3.29 Optional Equipment Enclosure Installation

If the optional enclosure is used, it may be configured to include other options,
including, but not limited to, a battery pack to provide reserve power to the NGC,
communication equipment, solar power charger and additional I/O.
Three enclosures are commonly used for the NGC8206 installations: the 6200,
6700 and 6800 enclosure. The unit may be mounted on a 2" pipe or mounted on a
flat surface such as a wall.
The battery and solar panel are packed and shipped separately from the
enclosure.
Before beginning, review the procedures and the materials required for the
installation; inspect all power cables where they terminate and the connector for
breakage.

3.29.1 6200 Optional Equipment Enclosure

The 6200 can accommodate the following equipment:
The 6200 installation will be for AC and 24 VDC sites. There is no battery-backed
option in this installation.
3.29.2 6700 Optional Equipment Enclosure
The 6700 optional equipment enclosure can accommodate the following:
2101510 Rev. AG
When all sample and vent tubing has been installed and
leak-tested, the sample boot must be sealed. The following
procedures describe the steps for sealing the environmental
enclosure.
insulating foam from inside the enclosure pointing toward the outside of
boot, ensuring that the overspray falls outside the enclosure.
The optional equipment enclosure may be approved for
classified hazardous locations or potentially explosive
atmospheres. Verify the rating listed on the unit tag and install
per the referenced control drawing. Be sure to follow the
requirements of national and local codes when installing the
optional equipment enclosure.
Power Supply Kit for 6200
120 VAC/12 VDC power supply
240 VAC/12 VDC power supply
Power Supply Kit
120/240 VAC / 12 VDC power supply
24 VDC/12 VDC DC to DC converter
ABB Totalflow XFC/XRC electronic board
XFC/XRC onboard battery charger
Battery Options
Page 3–55