4.2.2

TCP/IP Local Connection

Material Required:
4.2.2.1
1) Make local connection to the NGC using either a USB or RS-232 cable to
End View
1
8
Connector Clip
Not Used
8
Not Used
7
RX-
6
Not Used
5
Not Used
4
RX+
3
TX-
2
TX+
1
2) From the Analyzer Operation screen in PCCU, click Show Tree View in the
upper left corner of screen.
3) Click Communications to show the communication setup screen.
4) Select the Network tab. Enable DHCP. Send changes, and record a new IP
address for later use.
5) Exit PCCU, and disconnect the local communication cable.
6) Connect the Ethernet cross-over cable between the PC and NGC.
7) Open PCCU software. Click Operate on the menu bar. Navigate through the
drop-down list to Setup. From the fly-out menu, select System Setup.
8) Under Communications, set the PCCU com port to TCP, and enter the IP
address previously noted in the network ID or IP box. Close the System Setup
screen.
9) Verify the TCP/IP communications by clicking the Entry button on the main
screen (upper most left button).
Page 4–4
Ethernet Cross-Over Cable (see
Instructions
set up the initial parameters.

Figure 4-3 Ethernet Cable-Cross-Over

Figure
4-3).
Crossed-Over Cable
Brown
8
Brown/White
7
Green
6
Blue/White
5
Blue
4
Green/White
3
Orange
2
Orange/White
1
Brown
8
Brown/White
7
Orange
6
Blue/White
5
Blue
4
Orange/White
3
Green
2
Green/White
1
1
TX+
2
TX-
3
RX+
4
Not Used
5
Not Used
6
RX-
7
Not Used
8
Not Used
2101510 Rev. AG