Advertisement

5) Verify that the peaks are correctly labeled and integrated. If the peaks are
correctly labeled and integrated, return the unit to operation.
6) Allow unit to cycle 3-4 times.
7) Following the
ensuring that the Next Mode is set to Hold.
8) Reset the Alarm Enable to Yes. Verify that the alarm threshold is a valid
configuration.
9) Return the unit to regular operation.

6.3.13 Enclosure Temperature Alarm

If the Enclosure Temperature alarm is in warning status, the following procedure
will step the user through the troubleshooting process. On occasion, these
instructions may detour the user to other procedures, and, when complete, they
should return to these procedures to continue.
6.3.13.1
These alarms are indicative of either extremely high or low temperatures inside
the enclosure. Causes could range from external temperatures being extremely
high or low, to a bad temperature sensor on the analytical board.
6.3.13.2
1) Compare the outside temperature with the temperature reading on the Analyzer
Operation screen, Enclosure Temperature. Atmospheric temperature could be
less than the enclosure temperature by as much as 20 degrees.
2) If the temperature differential does not seem reasonable, the analytical processor
assembly may have a bad temperature sensor. As this alarm is only a warning, it
will not effect the operation of the unit. The user may replace the analytical
module, as needed.

6.3.14 Power Supply Alarm

If the Power Supply alarm is in warning status, the following procedure will step
the user through the troubleshooting process. On occasion, these instructions may
detour the user to other procedures, and, when complete, they should return to
these procedures to continue.
6.3.14.1
These alarms are indicative of input voltage either below 11 volts or above 16
volts. Causes may range from a power supply issue to a bad cable.
2101510 Rev. AG
Calibrating the NGC
The information provided for troubleshooting this alarm is only
intended to cover basic steps that can be performed in the
field. On occasion, additional troubleshooting steps may be
provided by Totalflow technical support in an effort to reduce
down time. Additionally, it may be desirable to return a
module to Totalflow for comprehensive testing and/or repair.
Description
Instructions
If the temperature differential seems reasonable, the unit may be operating
out of range. This unit is designed to operate between 08F and 1208F.
The Totalflow repair department offers a range of services for
troubleshooting and repairing/replacing the non-functioning
parts. For more information regarding the repair service,
contact customer service:
USA: (800) 442-3097 or International: 1-918-338-4880
Description
instructions in Startup, perform a calibration
Page 6–15