Operation
Calibration
3-16
42iQHL Instruction Manual
Home Screen>Calibration>Calibrate Span Coefficients>Calibrate High
Range NO Span Coefficient (dual or auto range mode)
The Calibrate NO, NO
following information:
Edit NO Span Concentration: User enters the NO span concentration
when in single range mode.
Edit Span Concentration: User enters the high range or low range span
concentration when in dual or auto range mode.
Current NO Concentration: Read only. Current NO concentration
reading when in single range mode.
Current High Range NO Concentration: Read only. Current high range
NO concentration reading when in dual or auto range mode.
Current Low Range NO Concentration: Read only. Current low range
NO concentration reading when in dual or auto range mode.
Current NO Span Coefficient: Read only. Displays the current user-set
NO span coefficient when in single range mode.
Current High Range NO Coefficient: Read only. Displays the current
user-set high range NO span coefficient when in dual or auto range
mode.
Current Low Range NO Coefficient: Read only. Displays the current
user-set low range NO span coefficient when in dual or auto range
mode.
Calculated NO Span Coefficient: Read only. After the "Edit NO Span
Concentration" value is entered, the new calculated NO span
coefficient is displayed.
Calculated High Range NO Span Coefficient: Read only. After the "Edit
Span Concentration" value is entered, the new calculated high range
NO span coefficient is displayed.
, and NO
Span Coefficient screens contain the
2
x
Thermo Scientific