Download Print this page

Specifications; Table 1-1. 42Iqhl Specifications - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Specifications

Thermo Scientific
Table 1–1
lists the specifications for the 42iQHL.
Table 1–1. 42iQHL Specifications
Range
Zero Noise
Detection Limit
Zero Drift
Span Drift
Response Time (in manual
mode)
Vacuum
Linearity
Sample Flow Rate
Bypass Flow Rate (option)
Operating Temperature Range
Power Requirements
Physical Dimensions
Weight
Analog I/O
Digital I/O
Serial Ports
Other Ports
Communication Protocols
Approvals and Certifications
05000 ppm
3
07500 mg/m
0.025 ppm RMS (60 second averaging time)
0.050 ppm (60 second averaging time)
0.050 ppm (24 hour)
±1% full-scale
2.5 seconds NO
5.0 seconds NO
x
28,.5" Hg
±1% full-scale
25 cc/min.
250–1100 cc/min.
0–
C
40 °
100–240 VAC 50/60 Hz
275 Watts
24 in (D) x 16.75 in (W) x 8.72 in (H) [609 mm (D) 425.45 mm
(W) x 221.48 mm (H)]
60 lbs
4 Isolated Voltage Inputs 0–10 V
6 Isolated Analog Voltages Outputs, with 4 selectable
ranges
6 Isolated Analog Current Outputs, with 2 selectable ranges
16 Digital Inputs (TTL)
8 Solenoid Driver Outputs
10 Digital Reed Relay Contact Outputs
1 RS-232/485 port
1 RS-485 External Accessory port
3 Full Speed USB ports (one in front, two in rear)
1 Gigabit Ethernet port
MODBUS, Bayern Hessen, Streaming
CE, TUV-SUD Safety