Calibrate No, No - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Operation
Calibration

Calibrate NO, NO

, and
The Calibrate NO, NO
, and NO
Span Coefficient screens are used to
2
2
x
enter span concentrations and calibrate the NO, NO
, and NO
NO
Span Coefficient
2
x
x
coefficients. The NO span coefficient is calculated, stored, and used to
correct the current reading. All calibration screens function the same way.
Therefore, the following example of the NO screen applies to the NO
and
2
NO
calibration screens as well.
x
The following screens are shown in single range mode and dual or auto
range mode. In dual or auto range modes, "High" or "Low" is displayed to
indicate the calibration of the high or low coefficient. The Calibrate High
Span Coefficient and Calibrate Low Span Coefficient screens function the
same way.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate Span Coefficients>Calibrate NO
Span Coefficient (single range mode)
Thermo Scientific
42iQHL Instruction Manual
3-15