Manual Entry: Shows current value. If the PMT supply cell is
pressed, user can manually change the value.
Target Concentration: Displays target concentration. If target
concentration cell is pressed, user can enter target concentration.
Calibration Concentration: Displays calibration concentration.
Auto Calibration: When pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
PMT voltage is automatically adjusted until calibration
concentration is with 1% of target concentration.
End Calibration: When pressed, the auto calibration is interrupted
and the value does not change.
Default PMT Supply: When pressed, the default PMT supply value
is saved.