Operation
Settings
NO
3-56
42iQHL Instruction Manual
Converter
The NO
2
converter. If an item being monitored goes outside the lower or higher
alarm limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>NO
Converter
The NO
Across:
Down:
Converter screen provides status and alarms related to the NO
2
Converter screen contains the following information:
2
Converter: This column lists items associated with the NO
converter.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
NO
Converter Temperature: Displays the current NO
2
temperature reading. User can adjust low and high alarm limits.
NO
Thermocouple Open: Displays OK/Fail for NO
2
open.
NO
Thermocouple Short: Displays OK/Fail for NO
2
short.
Board Communication: Displays OK/Fail for board communication
status.
2
2
2
converter
2
thermocouple
2
thermocouple
2