Operation
Settings
Flow and Pressure
3-60
42iQHL Instruction Manual
The Flow and Pressure screen provides status and alarms related to the flow
and pressure module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Flow and
Pressure
The Flow and Pressure screen contains the following information:
Across:
Flow and Pressure: This column lists items associated with the flow
and pressure module.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
Down:
Flow: Displays the current flow reading. User can adjust low and
high alarm limits.
Chamber Pressure: Displays the current chamber pressure reading.
User can adjust low and high alarm limits.
Sample Input Pressure: Displays the current sample input pressure
reading.
Instrument Temperature: Displays the current instrument
temperature reading. User can adjust low and high alarm limits.
Thermo Scientific