Calibration - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Calibration

Thermo Scientific
The Calibration screen allows the user to calibrate the system, setup
automatic calibrations, and view calibration data. See Chapter 4
"Calibration" for further instructions on how to run a calibration.
Home Screen>Calibration (with O
The Calibration screen contains the following information:
Calibrate Backgrounds: Sets the NO and NO
Calibrate Span Coefficients: Sets the NO, NO
coefficients.
Zero/Span Schedule: Programs the instrument to perform fully
automated zero and span checks or adjustments.
Calibrate O
Span Coefficient: When the O
2
(in the Settings>Configuration screen), this button will appear. It
allows an O
calibration to be performed.
2
Advanced Calibration: Calibrates the instrument using manual
zero/span calibration and provides calibration history.
option)
2
reading to zero.
x
, and NO
2
Sensor option is turned on
2
42iQHL Instruction Manual
Operation
Calibration
span
x
3-9