DMC Ozonator

Ozonator and
Transformer
Thermo Scientific
Use the following procedure to replace the ozonator and transformer board,
replace the flow switch and optional ammonia scrubber.
Removing the ozonator and transformer
Replacing the ozonator and transformer board
Replacing the flow switch
Replacing the ammonia scrubber (optional)
Use the following procedure to remove the ozonator and transformer.
Equipment Required:
Removal
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover
2. Unplug converter DMC (J2) and flow pressure board (J4).
Figure 7–46. HVPS, Electrical
3. Disconnect plumbing.
(Figure
2–1).
Servicing
DMC Ozonator
42iQHL Instruction Manual
7-45