Reaction Chamber Cleaning And/Or Removal; Figure 7-36. Photo Reaction Chamber Cleaning, Electrical - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Reaction Chamber
Cleaning and/or
Thermo Scientific
Use the following procedure to clean or remove the reaction chamber.
Equipment Required:
Removal
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove DMC Converter as described in
Removal" in this chapter.
3. Unplug heater J8 from board.
Figure 7–36. Photo Reaction Chamber Cleaning, Electrical
4. Disconnect plumbing as shown.
5. Using a 9/64-inch hex drive, unfasten three #8-32 socket cap head
screws.
DMC PMT Cooler and Reaction Chamber Replacement
"PMT Cooler Shroud
42iQHL Instruction Manual
Servicing
7-35