Servicing
Solenoid Valve Removal
3. Using a #1 Phillips drive, loosen two #4-40 pan head screws.
4. Slide valve out.
Equipment Damage Do not disconnect the plumbing from the valve end.
Disconnect from the attaching end only. This will prevent damaging and
leaks from the valve end.
Figure 7–57. Mode Hardware, Plumbing
Thermo Scientific
42iQHL Instruction Manual
7-55