Thermo Scientific 42iQHL Instruction Manual page 203

High level no-no2-nox analyzer

Advertisement

Thermo Scientific
Problem
Non-linear response
Excessive response time
Improper converter
operation
Alarm – Instrument
Temperature
Alarm – Bench
Temperature
Alarm – Cooler Temp
Sensor
Alarm – NO
Converter
2
Temperature
Alarm – Pressure
Alarm – Flow
Possible Cause
Action
Defective cooler
Check temperature (less than -2
°C at T
Incorrect calibration
Verify accuracy of multipoint
source
calibration source gas.
Leak in sample probe line
Check for variable dilution.
Partially blocked sample
Unplug power cord. Clean or
capillary
replace capillary.
Hang up/blockage in
Change element.
sample filter
Questionable calibration
Verify accuracy.
gas
Converter temperature
Temperature should be 625 °C for
too high or too low
stainless steel.
Low line voltage
Check to see if line voltage is
within specifications.
Fan failure
Replace fan if not operating
properly.
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting
the error and replace if needed.
Heaters failed
Replace heaters as needed.
Defective PCB
Replace PCB.
Check fan operation
Replace defective fan.
Check fan filter
Clean or replace filter.
Bad thermistor
Replace thermistor.
Converter temperature
Stainless steel converter should
low
be hot to the touch. If not, the
heater may have failed. Check
that converter temperature
setpoint is approximately 625 °C.
High pressure indication
Check plumbing for leaks.
Check the pump for a tear in the
diaphragm.
Replace if needed.
Check the capillary is properly
installed and o-rings are in good
shape.
Flow low
Check sample capillary for
blockage. Replace as necessary.
Troubleshooting
Troubleshooting Guide
= 25 °C).
amb
42iQHL Instruction Manual
6-3