Calibrate No X High; Calibrate No 2 High - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Calibrate NO
Calibrate NO
Thermo Scientific
2. After the responses stabilize, from the Home screen, choose
Calibration>Calibrate Span Coefficients>Calibrate High Range NO
Span Coefficient.
The user sets the span concentration by pressing the Edit Span
Concentration button. The New Span Coefficient button will show
what the span coefficient will be set to if the Calibrate button is
pressed.
3. Enter the NO calibration gas concentration using the pushbuttons, and
then press Calibrate to calculate and save the new high range NO span
coefficient based on the entered span coefficient.
High
Use the following procedure to calibrate the NO
x
calibration gas.
1. Press the Back button to return to the Calibration menu and choose
Calibrate High Range NO
The user sets the span concentration by pressing the Edit Span
Concentration button. The New Span Coefficient button will show
what the span coefficient will be set to if the Calibrate button is
pressed.
2. Verify that the high NO
the high NO calibration gas concentration plus any known NO
impurity.
3. Enter the NO
and then press Calibrate to calculate and save the new high range NO
span coefficient based on the entered span coefficient.
High
Use the following procedure to calibrate the NO
2
calibration gas.
1. Adjust the O
produce a decrease in the high NO concentration equivalent to about
80% of the URL of the high NO
90% of the high NO concentration determined in the
High" procedure.
Span Coefficient.
x
calibration gas concentration is the same as
x
calibration gas concentration using the pushbuttons,
x
generator in the GPT system to generate sufficient O
3
range. The decrease must not exceed
2
Calibration in Dual Range and Auto Range Mode
channel to the NO
x
channel to the NO
2
"Calibrate NOx
42iQHL Instruction Manual
Calibration
x
2
x
2
to
3
4-19