DMC Ozonator
Ozonator and
Transformer Board
Replacement
7-48
42iQHL Instruction Manual
Figure 7–49. Replace Transformer
Use the following procedure to replace the ozonator and transformer board.
Equipment Required:
Phillips drive, #2
1. Remove DMC Ozonator as described in
chapter.
2. Unplug all electrical from HPVS board, ozonator (J1), flow switch (J5)
converter DMC (J2) and flow pressure board (J4).
3. Using a #2 Phillips drive, unfasten four #6-32 pan head screws.