Signal
Signal name
READY
Ready
RUN
Run screen
ERR
Error
BANKOUT1_A to
Current bank number This outputs the currently specified bank number.
E
HOLDSTAT1
Hold execution status Turns ON when the displacement sensor is in the hold sampling period.
RESETSTAT1
Reset execution state Turns ON when the displacement sensor is in the reset execution state.
LIGHT1
Logical beam lighting
state
STABILITY1
Measurement
position
ENABLE1
Measurement state
GATE1
Data output
completed
OR1
Overall judgment
result
HIGH1_T1-4
HIGH output
PASS1_T1-4
PASS Output
LOW1_T1-4
LOW output
ZEROSTAT1_T1-4 Zero reset state
Command code
Command code
Response code
Response code
Response data
Response data
210
EtherNet/IP Connection
Function
Turns OFF when the displacement sensor cannot execute control commands.
Turns ON when the displacement sensor can execute control commands.
Turns ON when the displacement sensor is in the RUN mode.
Turns OFF when the displacement sensor is in the FUN mode.
Turns ON when a displacement sensor error is detected.
Turns OFF when the displacement sensor is normal.
After it turns ON, it never turns OFF until the error clear signal (ERCLR signal) from the
user (Controller) turns ON.
It expresses the bank number in combinations of BANKOUTx_A to E.
(For details of combinations, see Reference.)
Turns OFF when the displacement sensor is outside the hold sampling period.
Turns OFF when the displacement sensor is in the reset non-execution state.
Turns ON when the logical beam is lit.
Turns OFF when the logical beam is out.
Turns ON when the measured value is in the measuring range.
Turns OFF when the measured value is outside the measuring range.
Turns ON when the displacement sensor is ready for measurement.
Turns OFF when the displacement sensor cannot measure (excessive or insufficient
received light, outside measuring range, Calibration ROM not mounted, during FUN mode
non-measurement).
Turns ON when the displacement sensor completes control data output when hold is set.
The displacement sensor automatically turns OFF one Gate period after turning ON.
Turns ON when even one of the judgment result of the displacement sensor TASK1 to 4 is
other than PASS.
Turns OFF when all of the judgment result of the displacement sensor TASK1 to 4 is
PASS.
Turns ON when the judgment result of the displacement sensor TASK1 to 4 is HIGH
(HIGH threshold < measured value).
Turns OFF when the judgment result of the displacement sensor TASK1 to 4 is other than
HIGH.
Turns ON when the judgment result of the displacement sensor TASK1 to 4 is PASS (LOW
threshold  measured value  HIGH threshold).
Turns OFF when the judgment result of the displacement sensor TASK1 to 4 is other than
PASS.
Turns ON when the judgment result of the displacement sensor TASK1 to 4 is LOW (LOW
threshold > measured value).
Turns OFF when the judgment result of the displacement sensor TASK1 to 4 is other than
LOW.
Turns ON when the displacement sensor TASK1 to 4 is in the zero reset execution state.
Turns OFF when the displacement sensor TASK1 to 4 is in the zero reset non-execution
state.
The executed command code is returned.
The response code of the executed command is stored.
The response data of the executed command is stored.