Messages Without Error Code - Smiths detection HI-SCAN 5170-A User Manual

Conventional x-ray inspection units hi-scan with hitrax electronics
Table of Contents

Advertisement

Error messages
Error code: 3502
The internal hard disk is faulty! System failures may
be expected! Please inform the supervisor!
Error code: 3503
After an error the data system has been automatically
repaired! Have all software settings checked or re-
stored and have a basic calibration carried out if ne-
cessary!
Error code: 4001
Radioactivity detector is not activated!

Messages without error code

Messages in alphabetical order
Date for next preventive maintenance reached!
Invalid card!
Last image transmission still in progress! Repeat
transmission?
95587412 01/01/09 © Smiths Heimann
How to correct operating troubles
Cause/Correction
 Functions as e.g. the IMS system*, TIP system*
possibly do not work correctly. The electronics
should be exchanged.
 Have a supervisor and your service center check
the software settings and have carried out a basic
calibration. Incorrectly separating the unit from
the mains can be the cause of the error. See also
the instructions of error code 3500.
 Ask a higher access user to activate the radio-
activity detector.
Cause/Correction
 Inform the personnel responsible for the main-
tenance of the X-ray inspection unit.
 The inserted card is no ISIS card or it is not pro-
grammed. Have the card checked.
 Wait a few seconds until you start a further trans-
mission attempt with
S
Attention!
If you cancel the transmission with
image will be deleted and cannot be used for
the recheck!
r
.
s
, the
1200
1-103