Ground unit (GNDU)

Basic characteristics

The ground unit (see
measurement ground signal shared by all installed 4200-SCS instrumentation. In addition, the
GNDU SENSE terminal provides access to the SMU SENSE LO signals.
Figure 1-26
Ground unit
Basic ground unit characteristics are summarized in

Basic circuit configurations

Ground unit connections

Figure 1-27
GNDU FORCE signal is circuit COMMON. The GNDU SENSE terminal is connected to each SMU
SENSE LO signal through a unique auto-sense resistor.
When the GNDU SENSE signal is connected to a DUT, all measurements will be made relative to
this DUT connection.
1-32
Figure
1-26) provides convenient access to circuit COMMON, which is the
S
E
N
S
E
F
O
R
C
E
C
O
M
M
O
N
Table 1-4
Basic ground unit characteristics
Characteristic
Maximum current (FORCE triax connector)
Maximum current (COMMON binding post connector)
Maximum FORCE path/cable resistance
Maximum SENSE path/cable resistance
shows how the various GNDU signals are related to the SMU signals. Note that the
Return to
GNDU
Table
Section Topics
Model 4200-SCS User's Manual
1-4.
Description
2.6 A
9 A
1 Ω
10 Ω
4200-900-01 Rev. K / February 2017