Keithley 4200-SCS User Manual page 225

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Model 4200-SCS User's Manual
Table 3-29
Interconnect parts for 4200-SCS FLASH package
Quantity
6
4
4
4
2
4
4
4
Table 3-30
Tools supplied with the 4200-SCS FLASH package
Quantity
1
NOTE
To remove the LEMO triax-to-SMA adapter from a SMU, pull on the knurled silver
portion of the connector to release the latches and permit the adapter to separate
from the SMU connector.
CAUTION
The connection instructions below assume a four channel 4200-SCS FLASH
system, consisting of two 4205-PG2 cards (4 pulse channels) as well as four
SMUs, either 4200-SCS SMU or 4210-SMU, with SMU preamps removed (see
Figure 3-83
To test on-wafer devices, there are various ways to connect the supplied SMA
cables to the probe manipulators. For the direct connect method (see
and
cabling to the Triax connector compatible with many types of probe manipulators.
4200-900-01 Rev. K / February 2017
Description
SMA Tee, female – male – female
LEMO triax to SMA adapter
3 slot male triax to female BNC adapter
SMA male to BNC female adapter
4.25 in (10.8) cm white SMA cables
8 in (20.3 cm) white SMA cables
6.6 ft (2 m) white SMA cables
5 foot/ 1.5 m BNC cable
Description
SMA Torque wrench, 8 in-pound, with 5/16 head installed
Use the supplied torque wrench to tighten each connection as it is assembled.
Always connect and torque adapter/cable assemblies before attaching the assembly
to the instrument cards.
Pre-torquing eliminates any non-axial stress on the bulkhead connectors on the SMU
or pulse cards, which could cause damage to the cards installed in the 4200-SCS
chassis, requiring repair.
Failure to fully disengage the LEMO adapter latches may result in damage
to the adapter and/or the SMU, requiring repair.
or
Figure
3-84).
Figure
3-96) or switch method (see
Return to
Section 3: Common Device Characterization Tests
Figure
Section Topics
Comment
Trigger, combine SMU and PG2
channels
Adapt SMU Force output to SMA for
signal interconnect
Convert BNC cabling to Triax for
prober or switch matrix connection
Adapt Tee to BNC for cabling from
instrument to probe manipulators
Interconnect for triggering
Interconnect between PG2 and SMU
signals
Connect to probe manipulators
3-97), adapters convert the BNC
Figure 3-95
3-119