How To Generate Basic Pulses; Kpulse: Getting Started; Keithley Pulse Application (Kpulse); Starting Kpulse - Keithley 4200-SCS User Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 5: How to Generate Basic Pulses

KPulse: Getting started

Keithley Pulse Application (KPulse) is a graphical user interface (GUI) that is a non-programming
alternative to configure and control the installed Keithley pulse cards. It is used for quick tests
requiring minimal interaction with other 4200-SCS test resources.
The Kpulse application supports the source-only configuration of 4225-PMU, 4220-PGU, 4205-
PG2, and 4200-PG2 2-channel pulse cards. The 4225-PMU is identified as PMU on the card tab,
whereas all other card types are identified as VPU.

Starting KPulse

Open the KPulse GUI
example shows one PMU installed in the system.
From the GUI, each pulse generator can configure and control the following waveform types:
Standard pulse waveforms
run from the VPU or PMU tabs of KPulse. There is a tab for every Keithley pulse card
installed in the Model 4200-SCS.
Custom file arb waveforms
using the Arb Generator tab of KPulse. A VPU or PMU tab can then be used to load the
saved .kaf file into the pulse generator and run it.
Figure 5-1
KPulse GUI
5-2
How to Generate Basic PulsesKeithley Pulse Application (KPulse)
(Figure
5-1) by double-clicking the KPulse icon on the desktop. The GUI
and
Segment ARB
(full-arb): Pulses are configured and saved as a .kaf file
Return to
Section Topics
Model 4200-SCS User's Manual
waveforms: Pulses are configured and
4200-900-01 Rev. K / February 2017
Section 5

Advertisement

Table of Contents
loading

Table of Contents