Running Vds-Id-Pulse-Vs-Dc Utm - Keithley 4200-SCS User Manual

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Section 3: Common Device Characterization Tests
Figure 3-40
Default Definition tab and GUI For vds-id-pulse

Running vds-id-pulse-vs-dc UTM

The default settings are the same as the vds-id-pulse UTM, with the addition of
the DC measurement parameters (see
To run vds-id-pulse-vs-dc UTM:
1.
If measurement parameters (pulse average, NPLC, measure range) need to be set, use the
definition table.
2.
(Optional) If only source parameters need to be changed, use the UTM GUI by clicking the
GUI button on the vds-id-pulse-vs-DC test. Modify the source parameters in the GUI,
and click OK when finished.
3.
Click the green Run button. For a test with three curves and 40 points per curve, the test
should take about 1.5-2 minutes. During the test, neither the Graph tab or Sheet tab is
updated.
Figure 3-41
Default definition and typical graph for vds-id-pulse-vs-dc
3-42
Figure
Return to
Section Topics
Model 4200-SCS User's Manual
3-41):
4200-900-01 Rev. K / February 2017

Advertisement