Section 5: How to Generate Basic Pulses
Standard pulse waveforms
Standard pulse waveforms are configured and controlled from the pulse card tabs in the GUI.
Figure 5-3
Standard pulse waveform previewers: KPulse provides a preview of configured standard pulse
waveforms for each enabled channel. Each waveform previewer shows the high and low levels,
and timing for the waveform.
In
Figure
settings for KPulse (pulse high = 1 V and pulse low = 0 V). Channel 2 uses the same settings, but
the complement mode is enabled. Pulse high goes to pulse low level (0 V) and pulse low goes to
pulse high level (1 V).
5-4
explains how to use the GUI for Standard Pulse output.
5-3, the configuration shown in the waveform previewer for Channel 1 uses the default
Return to
Section Topics
Model 4200-SCS User's Manual
4200-900-01 Rev. K / February 2017