Reflect Mount - Siemens SITRANS FUH1010 Product Instruction Manual

Liquid ultrasonic flowmeter / interface detector (ifd)
Hide thumbs Also See for SITRANS FUH1010:
Table of Contents

Advertisement

Connecting
5.6 Sensor Installation
5.6.3

Reflect Mount

Mounting Frame Installation
1. On a flat surface, attach the Spacer Bar to a Mounting Frame so that the Reference Hole
2. Slide the second Mounting Frame onto the other end of the Spacer Bar and align the
3. Wrap a Mounting Strap around the pipe. Make sure to position it so there is easy access
4. At the mounting location, place the Mounting Frame/Spacer Bar Assembly on the pipe so
5. Engage the end of the Mounting Strap with the Mounting Strap Adjusting Screw.
6. Slide strap under the spring clip of one of the Mounting Frames.
7. Tighten the Mounting Strap Screw enough to take up all of the slack, but not enough to
8. Rotate the assembly on the pipe to the final conditioned location, ensuring that it is
9. Tighten the mounting straps to seat the assembly firmly on the pipe. Do not over tighten.
56
on the Spacer Bar fits over the metal post on the platform of the frame. Tighten the
clamping screw.
Number Index Hole with the metal post on the platform. Then tighten the clamping screw.
Ensure that the angled sides of both frames face away from each other.
to the Mounting Strap Adjusting Screw.
that it rests on the top of the pipe.
prevent rotation of the assembly. Attach the second mounting strap.
straight along the pipe axis. (Refer to the sensor orientation diagram)
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA