Figure 3-7
3.5

Installing sensors in Reflect mount

Reflect mount - Sensor installation using mounting frames and spacer bar
1. Use the Sensor settings Wizard setup procedure to program the meter for application
parameters such as pipe size, liquid type, and sensor size. Once entered the transmitter will
return a spacing index number appropriate for the application.
2. Using the supplied space bar, set the distance between the sensors according to the spacing
index number provided by transmitter. The spacing index provides the correct distance
between the sensors based on the pipe size given.
Before continuing refer to the Reflect mount installation diagram example below.
SITRANS FS230
Compact Operating Instructions, 05/2021, A5E38755607-AC
Mounting DSL to mount bracket
Installing/mounting
3.5 Installing sensors in Reflect mount