Non-medical grade displays
Isolation transformers
2026265-075C
Appendix A – Technical specifications: Solar 8000M/i display
The Solar 8000M/i patient monitor with a non-medical grade display
meets UL and IEC specifications if an isolation transformer is used,
regardless of whether the computer-grade display meets the leakage
current specification on its own.
The party assembling or modifying the medical electrical system is
responsible to insure compliance with IEC 60601-1-1. Therefore, if GE
installs a Solar 8000M/i system with a computer-grade display, GE is
responsible for meeting the specification.
As a result GE will only install non-medical grade displays with
appropriate isolation transformers. (See the following information.)
Powervar has designed an isolation transformer specifically for this
application. Powervar headquarters (listed below) will process orders
and drop-ship to any destination requested. When calling Powervar,
identify yourself as a GE representative/customer to receive the GE
partnership discount.