Setting Initial Operating Parameters - Waters 600e User Manual

Multisolvent delivery system
Hide thumbs Also See for 600e:
Table of Contents

Advertisement

Overview
The Waters 600E system treats a non-PowerLine autoinjector as an external device.
Programming Considerations
When you make injections using a non-PowerLine autoinjector through either the Direct
Control mode or Operate Method mode:
Specify the special vial code 100 in the Vial # field. This code tells the controller
that you are using an external injector. The inject start signal will be presented at the
Inject terminal on the rear panel.
If desired for documentation purposes, enter the injection volume in the Inj Vol
field.
When programming a method table, specify the number of injections.
Injecting Samples
To make sample injections using a non-PowerLine autoinjector:
1. Ensure that you have connected the injector according to the instructions in the
Waters 600E Multisolvent Delivery System Installation and Maintenance Guide.
2. Load the sample vials into the autoinjector.
3. Program the autoinjector as necessary.
4. Start the autoinjector when you are ready to make the run.

7.2 Setting Initial Operating Parameters

This section describes how to use the Operate Methods screens to set the initial operating
parameters for running a method table that you programmed (see
PowerLine
Methods).
Figure 7-2
summarizes the steps involved in using the Operate Methods screens. Use this
flowchart for quick reference once you become familiar with your Waters 600E system.