Waters 600e User Manual page 231

Multisolvent delivery system
Hide thumbs Also See for 600e:
Table of Contents

Advertisement

Table A-2 Electrical Specifications
Item
Power Requirements
Frequency
Fuse
Time- or Operator-
Controllable Switch
Closures
Auxiliary Power
Supply
Specification
Maximum power: 250 VA
47 to 63 Hz
Nominal
Voltage
Voltage
Range
100
10%
120
10%
230
10%
240
10%
Four controllable switch closures to ground as follows:
• TTL-compatible terminals pulled up with 10-K ohm resistors to 5 V.
May be pulled up to 12 V.
• Maximum allowable current = 1.0 A
• Maximum allowable voltage = 12 V. Outputs may be controlled
directly from keyboard or from events program.
• Two modes available from keyboard and Program Events screen:
ON = switch closed to ground
OFF = switch open, +5 V to ground
• Additional mode available from Program Events screen:
Pulse = single contact closure to ground for 600 msec
12 V to ground fused supply. 1 A max current.
Fuse
(North America)
2 A, 250 V
UL/CSA,
Time Delay
2 A, 250 V
UL/CSA,
Time Delay
1 A, 250 V
UL/CSA,
Time Delay
1 A, 250 V
UL/CSA,
Time Delay
Time Delay
Fuse
(Europe)
1.6, 250 V
IEC, Time
Delay
1.6, 250 V
IEC, Time
Delay
0.8, 250 V
IEC, Time
Delay
0.8, 250 V
IEC, Time
Delay
Delay
203