To modify a table that is currently running:
1. Stop the run by pressing the Abort screen key from the Operate Methods screen.
The Caution I screen appears.
2. Press the Abort screen key from the Caution I screen
3. Modify the table.
4. Save the table by pressing the Save screen key.

7.4.5 Modifying the Method Table

To modify any line in the method table that follows the currently running line:
1. Press the Program Method function key. The Program Methods screen appears.
2. Revise the table as necessary, then press the Save screen key.
3. Press the Operate Method function key to resume monitoring the run.

7.4.6 Aborting the Method

There are two ways to abort the method:
Pressing the Abort screen key
Pressing the Direct function key
Using the Abort Screen Key
To abort the method using the Abort screen key:
1. Press the Abort screen key. The Caution I screen appears