PowerLine Operation
The PowerLine controller provides product-specific setup and operating parameters for
PowerLine detectors and autoinjectors.
Detector parameters inlude:
Detector filter constant
Autozero (except for 431/432 detector)
Polarity
Absorbance Units Full Scale (AUFS)
UV wavelength (λ)
Sensitivity
Lamp On/Off
Filter
Cell temperature
Injector parameters include:
Syringe draw rate
Inject delay
Standards (how many and how often)
Methods specifying vial numbers
Number of injections from each vial
Injection volume
Storage for Tables
The controller provides storage for 15 sets of tables. Each set of tables can include a:
Gradient table - Contains a maximum of 15 operational steps that define the
varying eluent composition, flow rate, time range, and rate of change for each phase
of the gradient.
Event table - Contains a maximum of 15 operational steps to control output signals
to external devices, internal alarms, sparge rate, column heater temperature, detector
lamp On/Off, and linking operation to other tables.
Detector table - Contains a maximum of 15 operational steps to change wavelength
and attenuation during a run.
Comparing Controller Types
23