PowerLine Operation
The PowerLine controller provides product-specific setup and operating parameters for
PowerLine detectors and autoinjectors.
Detector parameters inlude:
•
Detector filter constant
•
Autozero (except for 431/432 detector)
•
Polarity
•
Absorbance Units Full Scale (AUFS)
•
UV wavelength (λ)
•
Sensitivity
•
Lamp On/Off
•
Filter
•
Cell temperature
Injector parameters include:
•
Syringe draw rate
•
Inject delay
•
Standards (how many and how often)
•
Methods specifying vial numbers
•
Number of injections from each vial
•
Injection volume
Storage for Tables
The controller provides storage for 15 sets of tables. Each set of tables can include a:
•
Gradient table - Contains a maximum of 15 operational steps that define the
varying eluent composition, flow rate, time range, and rate of change for each phase
of the gradient.
•
Event table - Contains a maximum of 15 operational steps to control output signals
to external devices, internal alarms, sparge rate, column heater temperature, detector
lamp On/Off, and linking operation to other tables.
•
Detector table - Contains a maximum of 15 operational steps to change wavelength
and attenuation during a run.