To set the Auto Standards parameter:
1. Move the cursor to the Auto Stds field.
2. Type 0 (No, default) to turn off auto standard sequencing, or type 1 (Yes) to select
auto standard sequencing. Press Enter.
3. If you typed 1 in step 2, enter values for the following parameters:
• Number of Standard Vials - Type the number of standard vials to run
(maximum 5) and press Enter.
• Sample Vials Between - Type the number of sample vials to be run between the
set of standards (1 to 48/96) and press Enter.
• Number of Injections - Type the number of injections to be made from each
standard vial (maximum 9) and press Enter.
• Injection Volume - Type the volume in µL (0 to 2000) for a standard injection
and press Enter.
Dissolution
Specifies whether you will be using a dissolution bath with your Waters WISP 712D
Autoinjector.
To set the Dissolution parameter:
1. Move the cursor to the Dissolution field.
2. Type 0 (No) or 1 (Yes).
3. Press Enter.
4.4.3 Exiting the Autoinjector Setup Screen
To exit the Autoinjector Setup screen:
•
Press the Next Setup screen key to display:
– The Detector Setup screen (if the value of the Dissolution parameter on the
Autoinjector Setup screen is No)
– The Dissolution Setup screen (if the value of the Dissolution parameter on the
Autoinjector Setup screen is Yes). See
Control.
•
Alternatively, press any function key to select another mode.
Section 4.5, Setting Up Dissolution
Setting Up the Autoinjector
75