To set the Auto Standards parameter:
1. Move the cursor to the Auto Stds field.
2. Type 0 (No, default) to turn off auto standard sequencing, or type 1 (Yes) to select
auto standard sequencing. Press Enter.
3. If you typed 1 in step 2, enter values for the following parameters:
• Number of Standard Vials - Type the number of standard vials to run
(maximum 5) and press Enter.
• Sample Vials Between - Type the number of sample vials to be run between the
set of standards (1 to 48/96) and press Enter.
• Number of Injections - Type the number of injections to be made from each
standard vial (maximum 9) and press Enter.
• Injection Volume - Type the volume in µL (0 to 2000) for a standard injection
and press Enter.
Dissolution
Specifies whether you will be using a dissolution bath with your Waters WISP 712D
Autoinjector.
To set the Dissolution parameter:
1. Move the cursor to the Dissolution field.
2. Type 0 (No) or 1 (Yes).
3. Press Enter.

4.4.3 Exiting the Autoinjector Setup Screen

To exit the Autoinjector Setup screen:
Press the Next Setup screen key to display:
– The Detector Setup screen (if the value of the Dissolution parameter on the
Autoinjector Setup screen is No)
– The Dissolution Setup screen (if the value of the Dissolution parameter on the
Autoinjector Setup screen is Yes). See
Control.
Alternatively, press any function key to select another mode.
Section 4.5, Setting Up Dissolution
Setting Up the Autoinjector
75