To modify a table during a run:
1. Press the Program Method function key to display the gradient table or the Program
Table function key to display the event table.
2. Move the cursor to the Table # field and type the number of the table to modify.
Press Enter.
Note: You cannot modify the table that is currently running.
3. Modify the table.
4. Press the Save screen key to save the new table.
Note: You cannot assign the number of the table that is currently running to the modified
table. To modify a table that is currently running, stop the run. See
the
Run.

11.4.5 Switching to Another Table

To switch to another gradient-event table set while a run is in progress:
1. Move the cursor to the Tables # field on the Operate Gradient screen.
2. Type the new table number and press Enter. The initial Operate Gradient screen (see
Figure
on the Current line.
The initial conditions of the newly selected table are displayed on the New line, but
are not in effect.
3. Review the flow and eluent conditions on the New line to determine if there is any
eluent or flow rate incompatibility with the current conditions.
4. If you selected the incorrect table, move the cursor to the Table # field in the New
line and enter another table number. Repeat step 3.
Otherwise, press the Operate Method function key to continue. The second Operate
Gradient screen appears (see
5. Press the Initial screen key to go to the initial conditions of the new table.
11-3) appears with the current gradient-/event-table flow conditions shown
Figure
11-4).
Modifying Conditions While Running 199