In addition to the parameters shown on the screen, the Direct Control screen displays the
screen keys described in
Table 5-2 Direct Control Screen Keys
Key
Function
Setup
Sends the operating parameters (displayed on the setup screens)
to the PowerLine modules. Toggles with Start Run and Abort.
Start Run
Starts the isocratic run and the run time clock. Toggles with
Setup and Abort.
Stop Flow
Stops the pump. To start the pump again, type in a value for
Flow and press Enter.
More
Displays a second set of screen keys.
Lamp On
Available after you press More. Turns on the 486 detector lamp.
Toggles with Lamp Off.
Lamp Off
Available after you press More. Turns off the 486 detector
lamp. Toggles with Lamp On.
Abort
Stops the run in progress. Toggles with Setup and Start Run.
Help
Displays the Hhelp for the Direct Control screen.
The lower-left corner of the Direct Control screen displays a system status flag. This flag
is similar to the screen keys except that it appears in reverse video. The flag indicates the
statuses described in
Table
Table 5-3 Direct Control System Statuses
Status
Function
Idle
No run in progress.
Setup
System is downloading the module operating parameters.
Initial
Module parameters have been downloaded; system is ready to
start a run.
Inject Wait
System is waiting for an inject signal from a contact closure at
the Inject terminal or a signal over the IEEE-488 bus to start
analysis.
Running
Analysis in progress.
Table
5-2.
5-3.
Performing a Direct Control Run
95