3. The autoinjector draws the sample and makes the injection. The autoinjector sends a
contact closure signal to the Inject input terminal on the rear panel of the controller.
Using the Start Run Screen Key with a Non-PowerLine Autoinjector
For methods development and for runs that are not time-sensitive, you may also use the
Start Run screen key to start a run with a non-PowerLine autoinjector.
To begin a run with an autoinjector using the Start Run key:
1. Load the sample vials into the autoinjector.
2. Program and start the autoinjector. See the autoinjector documentation.
3. Press the Start Run key as soon as possible after the injection occurs. Note that the
Start Run key does not initiate an inject switch contact closure.
Starting the Run Time Clock
When the controller receives an Inject signal from an injector, the system status changes to
and the run time clock begins counting at 0.00. The system does not accept
Running
another Inject signal until the run concludes or you abort the run.
Completing a Run
When a run completes, the system status changes from

5.2.5 Stopping the Pump

Stopping the Pump without Aborting the Run
To stop the pump without aborting the run do one of the following:
Press the Stop Flow screen key.
Type a value of 0.0 at the Flow field and press Enter.
Resuming Flow
To resume flow, type in a flow rate value at the Flow field and press Enter.
Note: These actions do not affect the eluent composition percentages.