2. Press the Pump Config screen key. The Gradient Controller Pump Configuration
screen appears
Figure 8-7 Gradient Controller Pump Configuration Screen
3. Move the cursor to the IEEE-488 field.
4. Type a number from 2 to 29 that is a unique address on the IEEE-488 network. Press
Enter.
Note: Ensure that the address is unique within your IEEE-488 network. Otherwise,
communications problems will result.
5. Press the Configure screen key to reset the system software and to enable your
IEEE-488 address selection. The controller reboots and returns to the Gradient
Controller Powerup screen.
If you are setting up your system for the first time or you want to change one or more
pump parameter values, proceed to
proceed to:
Chapter 9, Running Gradient Controller in Isocratic Mode
Chapter 10, Creating Gradient Controller Time-Based Tables
Chapter 11, Running Gradients
156
Setting Gradient Controller Operating Parameters
(Figure
8-7).
Section 8.2, Setting Up the 600E
Pump. Otherwise,