The cause of the increased deviation must be found and rectified. A zero recalibration as
previously described can then be retried.
If the operator deems this deviation to be acceptable, set P377 to 1 to invoke an initial
zero calibration. Further deviation limits are now based on this new initial zero.
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
This is the end of zero calibration. Proceed with zero, or span recalibration, or
Note:
return to RUN.

Initial Zero

Perform an initial zero if necessary when calibration is out of range message is shown.
During this step the progress of the process will be shown on the display with a progress
indicator in %.
Access P377 and enter EDIT mode
P377 Initial Zero
Enter 1 to start initial Zero
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
Press
Initial Zero Calibration in progress
Current Reading:
Calibration complete. Deviation
Press ENTER to accept value
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
Page 46
Milltronics BW500 and BW500/L - OPERATING INSTRUCTIONS 7ML19985DK05
e.g. zero calibration is accepted and dis-
551418
played as the current zero
E
0
e.g. the current zero
530560
%
#####
the zero count being calculated while cali-
bration is in progress
e.g. the deviation from the previous zero
0.00
e.g. the new zero count if accepted
551413
if unacceptable, press
e.g. the current zero count
551413
to restart