Zero Calibration

To obtain an accurate and successful calibration, ensure that the required
Note:
criteria are met. Refer to
Press
Zero Calibration: Current Zero
Clear belt. Press ENTER to Start
Press
Initial Zero Calibration. In progress
Current Reading:
The duration of the Zero calibration is dependent on speed (P014), length (P016), and
revolutions (P360) of the belt.
Calibration Complete. Deviation
Press ENTER to accept value:
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to Start
Accepting the Zero returns to start of Zero. A new Zero can be performed, or continue to
Span
.
The moisture meter is ignored during calibration. If inclinometer is used, then
Note:
calibration is adjusted based on incline angle.
Span Calibration
When performing a Span Calibration where the test reference is ECal (P002 = 3), the
supplied test weight or test chain must not be applied, and the conveyor must be run
empty.
To obtain an accurate and successful calibration, ensure that the required
Note:
criteria are met. Refer to
With the conveyor stopped and locked out, apply the test weight or chain to the scale as
instructed in the associate manuals; then start the conveyor.
Page 38
Milltronics BW500 and BW500/L - OPERATING INSTRUCTIONS 7ML19985DK05
Calibration Criteria
on page 155.
0
%
#####
0.00
551205
551205
Calibration Criteria
on page 155.
the current zero count
the zero count being calculated while cali-
bration is in progress
the deviation from previous zero. For an
initial zero there is no previous zero; hence
the deviation is 0.
e.g. the new zero count, if accepted