Each time one of the eight conditions is in effect, the corresponding multispan is selected
prior to putting the BW500 into the RUN mode. The selection is made by either changing
the multispan operation number, accessed via P365, or by external contacts connected to
the Auxiliary input, and programmed via P270.
To enable multispan operation, the following must be addressed.
connections
programming
calibration
operation
Connections
If the span selection is to be done by remote contact, the following connections would
apply. Otherwise, no additional connections to the BW500 are required.
Multispan Selection of Spans 1 and 2
*Remote contact can be from relay, open collector switch, or BCD switch.
Programming
Access P365 and enter EDIT mode
P365 Multispan
Select [1-8]
Span 1 will have already been set as part of the Start Up and initial calibration. Therefore,
select 2.
Access P017 and enter EDIT mode
P017 Test Load: Weight MS2
Enter test load
Enter the test load value, and press
7ML19985DK05Milltronics BW500 and BW500/L - OPERATING INSTRUCTIONS
Multispan Selection of Spans 1 to 8
E
0
E
0
to do a span calibration.