Setting Output Contact - YOKOGAWA ZR402G User Manual

Integrated type zirconia oxygen/humidity
Hide thumbs Also See for ZR402G:
Table of Contents

Advertisement

Table 8.9
Output contact 1
Output contact 2
8.4.2

Setting Output Contact

Set the output contacts following Table 8.10.
Table 8.10
Set item
Output contact 1
Operation
Error
High-high alarm
High alarm
Low alarm
Low-low alarm
Maintenance
Calibration
Measurement range
change
Warm-up
Calibration gas
pressure decrease
Unburnt gas
detection
Note 1:
Output contact 2 remains closed.
Note 2:
The oxygen concentration alarm must be preset (see Section 8.4).
Note 3:
Range change answer-back signal. For this action, the range change must be preset during the setting of input contacts
(see Section 8.6).
Note 4:
Calibration gas pressure decrease answer-back signal. Calibration gas pressure decrease must be selected beforehand
during the setting of input contacts.
Note 5:
Non-combusted gas detection answer-back signals. "Non-combusted gas" detection must be selected during the setting of
input contacts.
WARNING
Output contact 2 is linked to the detector's heater power safety switch. As such, if output
contact 2 is on, the heater power stops and an Error 1 (cell voltage abnormal) or Error 2 (heater
temperature abnormal) occurs.
Setting Output Contacts
Operating state
Open (de-energized) or closed
(energized) selectable.
Closed (de-energized) only.
Parameter Codes for Output Contact Setting
Parameter
code
E10
0 Operated in closed status. (Normally de-energized)
1 Operated when open. (Normally energized)
E20
0 Not operated if an error occurs.
1 Operated if an error occurs.
E21
0 Not operated if a high-high alarm occurs.
1 Operated if a high-high alarm occurs.
E22
0 Not operated if a high alarm occurs.
1 Operated if a high alarm occurs.
E23
0 Not operated if a low alarm occurs.
1 Operated if a low alarm occurs.
E24
0 Not operated if a low-low alarm occurs.
1 Operated if a low-low alarm occurs.
E25
0 Not operated during maintenance.
1 Operated during maintenance (see Section 8.3.1).
E26
0 Not operated during calibration.
1 Operated during calibration (see Section 8.3.1).
E27
0 Not operated when changing ranges.
1 Operated when changing ranges.
E28
0 Not operated during warming up.
1 Operated during warming up.
E29
0 Not operated while a calibration gas pressure decrease contact is
being closed.
1 Operated while a calibration gas pressure decrease contact is being
closed.
E32
0 Not operated while a unburnt gas detection contact is being closed.
1 Operated while a unburnt gas detection contact is being closed.
<8. Detailed Data Setting>
When no power is applied to this equipment
Set value
(Note 2)
(Note 2)
(Note 3)
(Note 4)
Open
Closed
(Note 1)
(Note 2)
(Note 2)
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
8-11
(Note 5)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22aZr202gExaxt zr seriesZr202aZr22g

Table of Contents