4.4.1
Piping Parts for Oxygen Analyzer with Pressure
Compensation
Check that the parts listed in Table 4.3 are provided.
Table 4.3
Equipment
Oxygen
Calibration gas inlet Check valve or stop
Analyzer
with pressure
compensation
Reference gas inlet
Note: Use parts with marking * as required.
General parts can be found on the local market.
4.4.2
Piping for the Calibration Gas
Calibration gas piping is basically identical to that of System 2. See Section 4.2.2.
4.4.3
Piping for the Reference Gas
Reference gas piping is basically identical to that of for System 2. See Section 4.2.3.
Piping Parts
Piping location
Parts Name
valve
Nipple *
Zero gas cylinder
Pressure regulator
Joint for tube connection Rc1/4 or 1/4 NPT
Air set
Joint for tube connection Rc1/4 or 1/4 NPT
<4. Piping>
Description
Stop valve (L9852CB or G7016XH)
recommended by YOKOGAWA
Check valve (K9292DN or K9292DS)
provided by YOKOGAWA
Rc1/4 or 1/4 NPT
User's scope
(G7013XF or G7014XF) recommended by
YOKOGAWA
(G7003XF/ K9473XK or G7004XF /
K9473XG) recommended by YOKOGAWA
IM 11M12A01-04E
4-8
General parts
General parts
General parts
11th Edition : Jul. 19, 2017-00