Linearity:
Drift:
Response Time:
Installation Altitude:
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage. Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
Degree 2 is the normal indoor environment.
Safety, EMC and RoHS conforming standards the ZR202G
Safety:
EMC:
Note:
RoHS:
2.1.2

ZR202G Integrated type Zirconia Oxygen Analyzer

Can be operated in the field without opening the cover using optical switches.
Display:
Switch:
Output Signal: 4 to 20 mA DC, one point (maximum load resistance 550 Ω)
Digital Communication (HART): 250 to 550 Ω, depending on quantity of field devices connected
Contact Output Signal: Two points (one is fail-safe, normally open)
Contact Input Signal:
(Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection)
(Use oxygen of known concentration (within the measuring range) as the
zero and span calibration gases.)
±1% Maximum value of set range;
less than 0 to 25 vol%O
(Sample gas pressure: within ±4.9 kPa)
±3% Maximum value of set range;
0 to 25 vol%O
(Sample gas pressure: within ±0.49 kPa)
±5% Maximum value of set range;
0 to 50 vol%O
(Sample gas pressure: within ±0.49 kPa)
In case of Humidity Analyzer,
± 2 vol% H
O; (Sample gas pressure: within ± 0.49 kPa)
2
± 3 vol% H
O; (Sample gas pressure: 2 kPa or less)
2
(Excluding the first two weeks in use)
(Excluding the case where the reference gas is by natural convection.)
Both zero and span ±2% Maximum value of set range/month
In case of Humidity Analyzer,
Both zero and span ± 3 vol% H
Response of 90% within 5 seconds. (Measured after gas is introduced
from calibration gas inlet and analog output start changing.)
2000 m or less
EN 61010-1, EN 61010-2-030, CAN/CSA-C22.2 No. 61010-1,
UL Std. No. 61010-1
EN 61326-1 Class A*, Table 2,
EN 61326-2-3, EN 61000-3-2
*: Influence of immunity environment (Criteria A ): ±20% of F. S.
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
EN61326-1 Class A
Korea Electromagnetic Conformity Standard
This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this
instrument in the industrial environment only.
EN 50581
6-digit LCD
Three optical switches
to the loop (multi-drop mode).
Two points
<2. Specifications>
range
2
or more and less than 0 to 50 vol%O
2
or more and up to 0 to 100 vol%O
2
O/month
2
IM 11M12A01-04E
2-2
range
2
range
2
11th Edition : Jul. 19, 2017-00