Characteristic Data From A Sensor Measured During Calibration - YOKOGAWA ZR402G User Manual

Integrated type zirconia oxygen/humidity
Hide thumbs Also See for ZR402G:
Table of Contents

Advertisement

81.92
ez
e2
Calibration curve
before correction
Cell
electromotive
force, mV
e1
es
C
0
21.0
Span gas concentration
Zero correction ratio = (B/A) x 100 (%)
Span correction ratio = (C/A) x 100 (%)
Figure 9.5
Figure 9.6 shows a one-point calibration using only a span gas. In this case, only the cell
electromotive force for a span gas with oxygen concentration p1 is measured. The cell
electromotive force for the zero gas is carried over from a previous measurement to obtain the
calibration curve. The principle of calibration using only a span gas also applies to the one-point
calibration method using a zero gas only.
81.92
ez
Cell
electromotive
Calibration curve before
force, mV
correction
e1
es
C
0
21.0
Zero correction ratio = (B/A) x 100 (%)
Span correction ratio = (C/A) x 100 (%)
Figure 9.6
9.1.5
Characteristic Data from a Sensor Measured During
Calibration
During calibration, calibration data and sensor status data (listed below) are acquired. However,
if the calibration is not properly conducted (an error occurs in automatic or semi-automatic
calibration), these data are not collected in the current calibration.
These data can be observed using parameter codes "A20" to "A22", and "A50" to "A79". For
an explanation and the operating procedures of individual data, consult Section 10.1, "Detailed
Display."
(1) Record of span correction ratio
Recorded the past ten span correction ratios.
(2) Record of zero correction ratio
Recorded the past ten zero correction ratios.
Corrected calibration
curve (theoretical
calibration curve)
p1
Zero gas concentration
Oxygen concentration (vol%O
)
2
Correctable range: 100 ± 30%
Correctable range: 0 ± 18%
Calculation of a Two-point Calibration Curve and Correction Ratios Using
Zero and Span Gases
Previous
zero gas data
Corrected calibration curve
(theoretical calibration curve)
p1
Span gas concentration
Oxygen concentration (vol%O
2
Correctable range: 100 ± 30%
Correctable range: 0 ± 18%
Calculation of a One-point Calibration Curve and Correction Ratios Using a Span Gas
Zero origin
B
A
Span origin
p2
0.51
F9.2E.ai
Zero origin
B
A
Span origin
0.51
)
F9.3E.ai
<9. Calibration>
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00