7.4.2

Display Configuration

The parameter codes provided for the equipment are used to control the equipment display
panels (see below). By selecting appropriate parameter codes, you can conduct calibration and
set operation parameters. Figure 7.5 shows the configuration of display items. The parameter
codes are listed in groups of seven; which are briefly described in Table 7.2.
To enter parameters, you first need to enter the password, refer to See 7.4.3.
Touch the [ >] key and [ ENT] key at same time to revert to the main screen.
Basic panel display
Password entry display
Group A setup display
Group B setup display
Group C setup display
Group D setup display
Group E setup display
Group F setup display
Group G setup display
Figure 7.5
Table 7.2
Display
Basic panel
Password entry Enters the password for the parameter code selection display.
Group A setup
Group B setup
Group C setup
Group D setup
Group E setup
Group F setup
Group G setup
%
Display Configuration
Display Functions
Function and item to be set
Displays the oxygen concentration in normal operation, or displays the detector
heater temperature while warming up. In case of humidity analyzer, displays
the oxygen con. or moisture quantity, or mixture ratio in normal operation. If an
error or alarm arises, the corresponding error or alarm number appears.
Displays detailed data, such as the cell voltage or temperature.
Sets and performs calibration.
Sets analog output.
Sets an alarm.
Sets the input and output contacts.
Selects the type of equipment and sets the parameters for computation.
Performs the current-loop or contact checks.
Parameter code
selection display
F7.4E.ai
<7. Startup>
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
7-4